IMADA specializes in
force measurement

MENU

Product Search

Solution Search

FAVORITES

Penetrability Test for Vial Stoppers (Complies with the corresponding part of ISO 8871-5:2016)

  1. Force Measurement IMADA
  2. SOLUTIONS
  3. Compression Test
  4. Penetrability Test for Vial Stoppers (Complies with the corresponding part of ISO 8871-5:2016)

Description FAVORITES

Penetrability Test for Vial Stoppers (Complies with the corresponding part of ISO 8871-5:2016)

Puncture Force

ISO

Graph

Medical devices and pharmaceuticals are used for diagnosis, treatment, and prevention of human and animal diseases, which are closely related to human and animal health. Therefore, high level of safety is required and many of these elements are strictly controlled by standards and internal regulations. This video shows how to perform a penetrability test of vial stoppers complying with ISO 8871-5: 2016. IMADA's specialized attachments, jig to fix the vial, Male Lure Slip Adapter to attach a hypodermic needle with 0.8 mm outer diameter (specified in the ISO standard), are used for this test. The peak force required for penetrating the vial stopper (rubber closure) is measured. The optional graphing software shows the entire force transition until the vial stopper (rubber closure) is penetrated, which helps you analyze the suitability of the vial stopper (rubber closure). It is used for the quality evaluation and quality control of vial.

inquiry number

1L2003A

Present the “inquiry number” below when you contact us, for prompt assistance.

Relevant Industrial Standards

ISO 8871-5 [ 2016 ] [Functional requirements and testing] (corresponding part only)

Related sample

  • Vial stoppers
  • Elastomeric closure

Measurement Example Videos

Cautions

The lure fitting adapter in this configuration is MED-LFA-20N-MS, male lure slip type. Refer to the specification sheet for other options of the adapters.
When replacing the sample needles, remove the taper connector of the lure fitting adapter first from the force gauge to protect the sensor of the gauge from overloading. In the case the sum of the length of the needle and the height of the vial exceeds 185mm, the test stand is required to have long stroke option.
This video shows the testing image of objective sample. Some samples may not be suitable to measure with this configuration depending on their materials or shapes.
The example of product configuration is for your reference only. It varies depending on sample, measurement method and testing capacity.
The products used in this video may differ from the example of product configuration.
Please select the right product model for the product configuration. Refer to the specification of each product for the compatibility.

CONTACT US
Unit
Converter
Unit Converter

Convenient Unit/Torque/
Tension conversion

6 Significant Digits (*Significant digits are meaningful digits that do not include zeros, which are used to indicate the scale.)

Products

CLICK
HERE!
Attachments, Grips & Fixtures Attachments, Grips & Fixtures
Custom Made Custom Made